首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Enzyme-linked immunosorbent assay in analysis of deoxynivalenol: investigation of the impact of sample matrix on results accuracy
Authors:Zbynek Dzuman  Marta Vaclavikova  Ivana Polisenska  Zdenka Veprikova  Marie Fenclova  Milena Zachariasova  Jana Hajslova
Institution:1.ICT Prague,Prague,Czech Republic;2.Agrotest fyto Ltd,Kromeriz,Czech Republic
Abstract:Enzyme-linked immunosorbent assay (ELISA) represents a bioanalytical strategy frequently used for rapid screening of mycotoxin deoxynivalenol (DON) in cereals and derived products. Due to a considerable affinity of some anti-DON antibodies to structurally similar DON metabolites, such as DON-3-glucoside (DON-3-Glc) and 3-acetyl-DON (3-ADON), a significant overestimation of DON concentrations may occur. A validation study of six commercial DON-dedicated ELISA kits, namely Ridascreen DON, Ridascreen FAST, DON, DON EIA, AgraQuant DON Assay, Veratox 5/5, and Veratox HS was carried out on wheat, barley, and malt matrices. Performance characteristics of all tested ELISAs were determined using aqueous solutions of DON, DON-3-Glc, and 3-ADON analytical standards, further with extracts of artificially spiked blank cereals, and finally with matrix-matched standards of all three compounds. In the final phase, the accuracy of data was assessed through a comparison of DON concentrations determined by particular ELISAs and reference ultra-high-performance liquid chromatography–tandem mass spectrometry method. For this purpose, both quality control materials and a comprehensive set of naturally and artificially contaminated samples of wheat, barley, and malt were analyzed. High cross-reactivities were proved for both DON-3-Glc and 3-ADON in the majority of examined assays, and moreover, a considerable contribution of some matrix components to overestimation of DON results was confirmed.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号