首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of arsenic in biological fluids by electrothermal atomic absorption spectrometry
Authors:Campillo N  Viñas P  López-García I  Hernández-Córdoba M
Affiliation:Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, Spain.
Abstract:A procedure for the determination of the total content of arsenic in urine, serum and blood by electrothermal atomic absorption spectrometry (ETAAS) is described. Zeeman correction is used to compensate the high background signals. The samples are diluted (1 + 1 for urine and 1 + 3 for both serum and blood samples) in a medium containing 0.1% w/v Triton X-100 before being introduced directly into the furnace. A solution containing 15% w/v hydrogen peroxide, 0.65% w/v nitric acid and 0.5% w/v nickel is also introduced into the atomizer by means of a separate injection. Calibration is carried out against aqueous standards for blood and serum samples and using the standard additions method for urine samples. The detection limit is 20 pg (2 ng ml-1). The reliability of the procedure is checked by analyzing three certified reference materials and by recovery studies.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号