Abstract:A continuous ultrasound-assisted extractor was coupled to a photometric detector in order to obtain a fully automated approach for the determination of CrVI in soil. The use of a flow injection (FI) manifold as interface between the extractor and the photometric detector allowed the monitoring of CrVI after extraction in a continuous manner. The coloured complex formed between 1,5-diphenylcarbazide (DPC) and CrVI was used as recommended in EPA method 7196A because it is one of the most sensitive and selective reactions for CrVI determination. A preconcentration minicolumn packed with a strong anion-exchange resin was placed between the extractor and the detector, providing a more sensitive method. The linear dynamic ranges were 1-10 and 0.25-7.5 mg l-1 for the methods without (method A) and with preconcentration (method B), respectively. The limits of detection were 4.52 ng for method A and 1.23 ng for method B. Both methods were applied to a natural contaminated soil and the results obtained agreed well with those obtained by the reference EPA method 3060A. The influence of different amounts of CrIII in the samples was also studied and the results showed that the proposed methods did not disturb the original species distribution.