首页 | 本学科首页   官方微博 | 高级检索  
     


Enantioselective gas chromatographic analysis of aqueous samples by on-line derivatisation. Application to enzymatic reactions
Authors:Mommers John H M  de Wildeman Stefaan M A  Koolen Wilma A F  Duchateau Alexander L L
Affiliation:DSM Resolve, 6160 MD Geleen, The Netherlands. John.Mommers@dsm.com
Abstract:A new gas chromatography (GC) method is presented for analysing both the conversion and the enantiomeric excess (e.e.) of samples from alcohol dehydrogenase reactions. The chiral compounds studied were a series of saturated, straight chain alcohols, ranging from 2-butanol to 2-heptanol. The alcohols were converted to the corresponding trifluoroacetylated derivatives by injecting trifluoroacetic anhydride onto the column shortly after injection of the aqueous samples in split-injection mode (1:100) onto a Chiraldex G-TA capillary GC column. Injecting seven hundred aqueous enzymatic reaction mixtures according to the above-mentioned procedure revealed no noticeable loss of column performance. Using the new GC method, conventional sample work-up procedures such as extraction and off-line derivatisation are eliminated and throughput of samples is significantly enhanced.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号