首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Rapid protein identification using a disposable on-line clean-up/concentrating device and electrospray ionization mass spectrometry
Authors:Tsai Chu-Yun  Pai Pei-Jing  Ho Yi-Hsin  Lu Jyh-Feng  Wang Jinn-Shyan  Lin Wann-Yin  Her Guor-Rong
Institution:Department of Chemistry, National Taiwan University, Taipei 106, Taiwan, ROC.
Abstract:A simple, low-cost, expedient method has been developed for identification of proteins isolated from two-dimensional (2D) gels. The method described uses a disposable on-line clean-up device, a syringe infusion pump and electrospray ionization mass spectrometry (ESI-MS). The on-line clean-up and concentrating device is a tapered capillary column filled with 1.5 cm of 5 microm C18 particles. The short column was easily prepared and was connected directly to the ESI source through a low-flow ESI sprayer. Peptides resulting from enzymatic digestion of proteins were eluted from the short column isocratically using a syringe infusion pump and analyzed by ESI-MS. This simple set-up was found useful in the analysis of proteins isolated from 2D gels. Compared to the more conventional micro-liquid chromatography/tandem mass spectrometry (microLC/MS/MS), this method can identify proteins rapidly without the need for an HPLC pump and removes the problem of cross-contamination caused by system carryover. These advantages make the method described competitive with conventional LC/MS even though the latter method gives slightly expanded sequence coverage.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏