首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Spectrophotometric determination of urinary protein with o-sulfophenylfluorone-metal complex.
Authors:Takako Yamaguchi  Emi Amano  Shinichiro Kamino  Satoko Umehara  Chizuko Yanaihara  Yoshikazu Fujita
Institution:Osaka University of Pharmaceutical Sciences, 4-20-1, Nasahara, Takatsuki, Osaka, Japan.
Abstract:A simple and sensitive spectrophotometric method for the determination of human serum albumin (HSA) was established based on the ternary complex-formation reaction of HSA with o-sulfophenylfluorone (SPF) as a xanthene dye and metal ion (niobium(V) and bismuth(III)) in the presence of a dispersion agent. This new method enabled the determination of HSA in the range of 1 - 15 microg/ml HSA by measuring the difference of the absorbance at 530 nm between HSA-SPF-metal ion and SPF-metal ion solutions. In the determination of HSA, this method is about 2-times more sensitive than the Pyrogallol Red-molybdenum(VI) method (PR method), which accounts for more than 80% of the quantification methods for urinary protein assays in Japan. There was no significant difference between the results obtained by the present method and the PR method for human urine samples. The binding process between the SPF-metal complex and HSA was studied by determining the binding parameters and the thermodynamic parameters.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号