Higher-order cross-correlation-based Doppler optical coherence tomography
Authors:Huang Liangmin  Ding Zhihua  Hong Wei  Wang Chuan  Wu Tong
Institution:State Key Laboratory of Modern Optical Instrumentation, Zhejiang University, Hangzhou, China.
Abstract:A method based on higher-order cross-correlation is proposed to fetch the Doppler information on flow velocity within areas under low signal-to-noise ratio (SNR) by spectral domain optical coherence tomography. The proposed method is theoretically developed and validated by measurement of a moving mirror with known velocities. Standard deviations of flow velocities of the mirror under different SNRs are determined by the proposed method and compared with those by the modified phase-resolved method. Measurement of flowing particles within a glass capillary is also conducted, and Doppler flow velocity maps of the glass capillary are reconstructed by both methods. All experimental results demonstrate that the proposed method can significantly suppress noise, thus rendering it suitable for flow measurement under low SNR cases.