采用激光诱导击穿光谱技术测定合金钢中锰和硅的含量
引用本文:孙兰香,于海斌,辛勇,丛智博. 采用激光诱导击穿光谱技术测定合金钢中锰和硅的含量[J]. 光谱学与光谱分析, 2010, 30(12): 3186-3190. DOI: 10.3964/j.issn.1000-0593(2010)12-3186-05
作者姓名:孙兰香  于海斌  辛勇  丛智博
作者单位:1. 中国科学院沈阳自动化研究所工业信息学重点实验室,辽宁 沈阳 110016
2. 中国科学院研究生院,北京 100049
基金项目:国家(863计划)项目,中国科学院知识创新工程重要方向项目
摘    要:基于激光诱导击穿光谱(LIBS)技术定量分析多品种合金钢中锰(Mn)和硅(Si)的浓度,采用的校正方法为多元二次非线性函数定标方法。分析的钢种有普通合金钢、不锈钢、碳素工具钢等。由于不同钢种在基体组成上存在较大差异,而且钢的光谱还存在复杂的谱线重叠干扰,所以受基体效应影响严重。因此,对钢这类复杂化学体系物质的定量分析,若采用单谱线信息的校正方法会造成很多有用信息的丢失,而采用多变量的定标法会更充分利用光谱中的信息,更有效校正多基体组分的基体效应,进而提高LIBS定量分析的重复性和准确性。相比线性标准定标曲线法,测量结果的相对标准差从20%以上降低到10%以下,Mn元素测量结果的准确度提高了5倍多,Si元素提高了6倍多。

关 键 词:激光诱导击穿光谱  基体效应  多元二次非线性函数  合金钢  
收稿时间:2010-02-12

Quantitative Analysis of Mn and Si of Alloy Stseels by Laser-Induced Breakdown Spectroscopy
SUN Lan-xiang,YU Hai-bin,XIN Yong,CONG Zhi-bo. Quantitative Analysis of Mn and Si of Alloy Stseels by Laser-Induced Breakdown Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2010, 30(12): 3186-3190. DOI: 10.3964/j.issn.1000-0593(2010)12-3186-05
Authors:SUN Lan-xiang  YU Hai-bin  XIN Yong  CONG Zhi-bo
Affiliation:1. Key Laboratory of Industrial Informatics, Shenyang Institute of Automation, Chinese Academy of Sciences, Shenyang 110016, China2. Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:The concentration of Mn and Si in different kinds of steels was determined by laser-induced breakdown spectroscopy (LIBS). The multivariate quadratic nonlinear function was adopted for calibration. Samples including common alloy steels, stainless steels and carbon tool steels were analyzed. The matrix effect was serious because of large difference in compositions of different kinds of steels and strong line overlaps in steel spectra. Therefore, the common calibration methods that only use one analytical line to calibrate the complex chemical compositions of alloy steels will lose much information. The multivariate calibration methods, however, can utilize more information of spectra, successfully reduce the matrix effect and improve the measurement repeatability and accuracy of LIBS. Compared with the common calibration method based on one analytical line, the relative standard deviation was reduced from above 20% to below 10%, and the accuracy was increased by more than 5 times for Mn and more than 6 times for Si.
Keywords:Laser-induced breakdown spectroscopy  Matrix effect  Multivariate quadratic nonlinear function  Alloy steels   