首页 | 本学科首页   官方微博 | 高级检索  
     


Styrene oxide DNA adducts: quantitative determination using 31P monitoring
Authors:Michael?Edler,Norbert?Jakubowski,Michael?Linscheid  author-information"  >  author-information__contact u-icon-before"  >  mailto:m.linscheid@chemie.hu-berlin.de"   title="  m.linscheid@chemie.hu-berlin.de"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author
Affiliation:(1) Institute for Analytical Sciences (ISAS), P.O. Box 101352, 44013 Dortmund, Germany;(2) Department of Chemistry, Humboldt University Berlin, Brook-Taylor-Str. 2, 12489 Berlin, Germany
Abstract:The reaction of styrene oxide, a potential carcinogen in humans, with DNA constituents has been used to develop an improved method for quantification of DNA adducts. To enable monitoring of DNA adducts caused by xenobiotics at physiological relevant levels, a robust, reliable and powerful method based on monitoring of phosphorus in nucleotides is described. An efficient enzymatic digestion step and a sample-preconcentration procedure are essential, and enable separation of alkylated nucleotides from the large excess of native nucleotides. The adducts are detected by means of the phosphorus signal measured at mass m/z=31 with an inductively-coupled-plasma mass spectrometer. Bis(4-nitrophenyl)phosphate (BNPP) serves as internal standard for quantification of the adducts. The absolute limit of detection, 45 fmol, corresponds to detection of three modified nucleotides among 107 native nucleotides (the calculation is based on use of 50 mgrg calf thymus DNA). An adduct formation ratio at the DNA of 3.6 adducts per 1000 nucleotides was measured, which is 75% lower than for reaction with monomeric 2prime-deoxy-nucleotides. In addition, a substantial amount of phosphate adducts were detected, but in DNA the rate of phosphate formation was lower than with monomeric nucleotides. Most probably these adducts escaped unnoticed when 31P-post-labelling was employed.
Keywords:DNA adducts  Inductively-coupled-plasma mass spectrometry (ICP–  MS)  31P monitoring  Styrene oxide
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏