Magnetic resonance imaging of rigid polymers at elevated temperatures with SPRITE |
| |
Authors: | B. MacMillan M. Halse M. Schneider L. Fardy Y. H. Chui B. J. Balcom |
| |
Affiliation: | (1) Pathogen Genomics, The Wellcome Trust Sanger Institute, Wellcome Trust Genome Campus, Hinxton, Cambridge, CB10 1SA, UK |
| |
Abstract: | Magnetic resonance imaging has rarely been applied to rigid polymeric materials, due primarily to the strong dipolar coupling and short signal lifetimes inherent in these materials. SPRITE (single point ramped imaging withT 1 enhancement) (B. J. Balcom, R. P. MacGregor, S. D. Beyea, D. P. Green, R. L. Armstrong, T. W. Bremner: J. Magn. Reson. A123, 131–134, 1996) is particularly well suited to imaging solid materials. With SPRITE, the only requirement is thatT 2* be long enough so that the signal can be phase-encoded. The minimum phase encoding time is limited by the maximum gradient strength available and by the instrument deadtime. At present this is usually tens of microseconds and will only improve with refinements in technology. We have used the SPRITE sequence in conjunction with raising the sample temperature to obtain images of rigid polymers that have largely frustrated conventional imaging methods. This approach provides a straightforward and reliable method for imaging a class of samples that, up until now, have been very difficult to image. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|