Affiliation: | aDepartment of Chemistry, Faculty of Sciences, University of Balearic Islands, Carretera de Valldemossa km 7.5, E-07122 Palma de Mallorca, Spain bREQUIMTE, Serviço de Química-Física, Faculdade de Farmácia, Universidade do Porto, Rua Aníbal Cunha 164, 4099-030 Porto, Portugal |
Abstract: | In this paper, a flow-through optosensor based on diffuse reflectance spectroscopy and multivariate regression modelling is proposed for simultaneous multiresidue determination of trace level concentrations of nitro-substituted phenols. The optrode integrates on-line anion-exchange sorptive preconcentration and matrix removal with direct detection of sorbed species onto the sensing layer and chemometric deconvolution of overlapped spectra. After recording of the reflectometric spectrum, fast chemosensor regeneration is accomplished with an acidic methanolic eluent followed by a metered volume of alkaline solution. The full automation of the analytical procedure is ensured by flow-programming as executed via a software-controlled multisyringe pump and a set of multiposition and solenoid valves. Under the optimized chemical and physical variables, the 3σblank detection limits for 2-, and 4-nitrophenol and 2,4-dinitrophenol were 0.69, 0.42 and 0.37 μg L−1, respectively, for a sample loading volume of 4.0 mL with enrichment factors of 73, 74 and 81, respectively, and repeatability better than 3.0%. The analytical applicability of the optosensing system was asserted by the satisfactory results attained in its application to monitoring of nitrophenol derivatives in spiked environmental waters including highly saline matrices via the standard additions method. |