首页 | 本学科首页   官方微博 | 高级检索  
     


True coincidence summing correction determination for 214Bi principal gamma lines in NORM samples
Authors:Kh. Haddad
Affiliation:1. Protection & Safety Department, Atomic Energy Commission, P.O. Box 6091, Damascus, Syria
Abstract:The gamma lines 609.3 and 1,120.3 keV are two of the most intensive γ emissions of 214Bi, but they have serious true coincidence summing (TCS) effects due to the complex decay schemes with multi-cascading transitions. TCS effects cause inaccurate count rate and hence erroneous results. A simple and easy experimental method for determination of TCS correction of 214Bi gamma lines was developed in this work using naturally occurring radioactive material samples. Height efficiency and self attenuation corrections were determined as well. The developed method has been formulated theoretically and validated experimentally. The corrections problems were solved simply with neither additional standard source nor simulation skills.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏