首页 | 本学科首页   官方微博 | 高级检索  
     


Thorium and iodine memory effects in inductively-coupled plasma mass spectrometry
Authors:Assad Al-Ammar  E. Reitznerová  R. M. Barnes
Affiliation:Department of Chemistry, Lederle Graduate Research Center Tower, University of Massachusetts, 710 N. Pleasant Street, Amherst, MA 01003-9336, and University Research Institute for Analytical Chemistry, 85 N. Whitney Street, Amherst, MA 01002-1869, USA e-mail: mbarnes@chemumass.edu, US
Abstract:Thorium and iodine memory effects have been characterized experimentally for inductively-coupled plasma mass spectrometry by adding ammonia gas directly to the spray chamber and nebulizing aqueous ammonia sample solutions to assess analyte memory retention sites. Thorium memory effect originates from the tendency of an unidentified thorium compound to volatilize from the spray-chamber walls, and not from Th compound adsorption to nebulizer tubing. The mass spectrometer skimmer and sampler cones, ion optics, quadrupole, and other components are not responsible for the memory effect. Unlike that of thorium the iodine memory effect originates from adsorption of iodine compounds on nebulizer tubing surfaces and from volatilization of HI and I2 from the spray-chamber walls. Addition of ammonia sample solutions or ammonia gas directly to the spray chamber eliminated the Th and I memory effects in practical analyses, and blank levels were achieved after 2?min wash-outs. Quantitative recoveries were obtained for Th and I in reference materials.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号