首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of L‐trantinterol in rat plasma by using chiral liquid chromatography‐tandem mass spectrometry
Authors:Lijuan Jing  Kunjie Li  Fen Qin  Xintao Wang  Li Pan  Yanjuan Wang  Maosheng Cheng  Famei Li
Affiliation:1. School of Pharmacy, Shenyang Pharmaceutical University, , Shenyang, China;2. School of Pharmaceutical Engineering, Shenyang Pharmaceutical University, , Shenyang, China
Abstract:A simple, sensitive, and rapid method for determination of L‐trantinterol in rat plasma was developed for the first time by using LC coupled to MS/MS based on chiral stationary phase. A baseline separation of the enantiomers of trantinterol was achieved on a Chirobiotic V column, using a mixture of acetonitrile–methanol–ammonia–acetic acid (80:20:0.01:0.02, v/v/v/v) as the mobile phase. The detection was performed on a triple‐quadrupole tandem mass spectrometer by multiple reaction monitoring mode via ESI. The calibration curve was linear in concentration range from 0.270 to 108 ng/mL in plasma with the lower limit of quantification of 0.270 ng/mL. The intra‐ and interday precision (relative standard deviation) values were within 10.9% and the accuracy (relative error) was from 2.6 to 9.2% at all quality control levels. The method has been successfully applied to a study of L‐trantinterol pharmacokinetics in rats.
Keywords:Chiral solid phase  LC‐MS/MS  L‐Trantinterol  Pharmacokinetics