首页 | 本学科首页   官方微博 | 高级检索  
     


Ultra-sensitive determination of cadmium in rice and water by UV-vis spectrophotometry after single drop microextraction
Authors:Wen Xiaodong  Deng Qingwen  Guo Jie  Yang Shengchun
Affiliation:College of Pharmacy and Chemistry, Dali University, Dali, Yunnan 671000, China. wenxdong129@163.com
Abstract:In this work, a new method based on single drop microextraction (SDME) preconcentration using tetrachloromethane (CCl(4)) as extraction solvent was proposed for the spectrophotometric determination of cadmium in rice and water samples. The influence factors relevant to SDME, such as type and volume of extractant, stirring rate and time, dithizone concentration, pH, drop volume and instrumental conditions were studied systematically. Under the optimal conditions, the limit of detection (LOD) was 0.5 ng L(-1), with sensitivity enhancement factor (EF) of 128. The different maximum absorption wavelength caused by the different extraction acidity compared with some conventional works and the enhancement effect of acetone (dilution solvent) for the spectrophotometric determination were the two key factors of the high EF and sensitivity. The proposed method was applied to the determination of rice and water samples with satisfactory analytical results. The proposed method was simple, rapid, cost-efficient and sensitive.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号