首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of glutathione disulfide levels in biological samples using thiol-disulfide exchanging agent, dithiothreitol
Authors:Ates Burhan  Ercal Baris Can  Manda Kalyan  Abraham Linu  Ercal Nuran
Institution:Department of Chemistry, Missouri University of Science and Technology, 400 West 11th Street, Rolla, MO 65409, USA.
Abstract:A reverse-phase HPLC method incorporating dithiothreitol (DTT) reduction for quantitative determination of oxidized glutathione (GSSG) in biological samples is described here. This method is based on our previous enzymatic reduction technique that uses N-1-(pyrenyl) maleimide (NPM) as a derivatizing agent. In our earlier method, glutathione disulfide (GSSG) was measured by first reducing it to GSH with glutathione reductase (GR) in the presence of NADPH. However, this is a very costly and time-consuming technique. The method described here employs a common and inexpensive thiol-disulfide exchanging agent, DTT, for reduction of GSSG to GSH, followed by derivatization with NPM. The calibration curves are linear over a concentration range of 25-1250 nm (r(2) > 0.995). The coefficients of variations for intra-run precision and inter-run precision range from 0.49 to 5.10% with an accuracy range of 1.78-6.15%. The percentage of relative recovery ranges from 97.3 to 103.2%. This new method provides a simple, efficient, and cost-effective way of determining glutathione disulfide levels with a 2.5 nm limit of detection per 5 microL injection volume.
Keywords:GSH  GSSG  DTT  thiol–disulfide exchange  HPLC
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏