首页 | 本学科首页   官方微博 | 高级检索  
     


Multisyringe flow-injection system for total inorganic arsenic determination by hydride generation-atomic fluorescence spectrometry
Authors:N.V SemenovaL.O Leal,R FortezaV Cerdà  
Affiliation:a Department of Geography, Lomonosov State University, Moscow 119899, Russia
b Department of Chemistry, University of the Balearic Islands, Carretera de Valldemossa km.7.5, E-07071 Palma de Mallorca, Spain
Abstract:A software-controlled time-based multisyringe flow-injection system for total inorganic arsenic determination by hydride generation atomic fluorescence spectrometry (HGAFS) has been developed. By using a multisyringe burette coupled with one multiport selection valve, the time-based injection provides precise known volumes of sample, a reducing sodium tetrahydroborate solution and a pre-reducing solution which are dispensed into a gas-liquid separation cell. An argon flow delivers the arsine into the flame of an atomic fluorescence spectrometer. A hydrogen flow has been used to support the flame.Linear calibration graphs for arsenic concentrations between 0.25 and 12 μg l−1 were obtained. The detection limit of the proposed technique (3σb/S) was 0.07 μg l−1. A sample throughput of 36 samples/h (108 injections) has been achieved. The proposed technique has been validated by means of reference solid and water materials with good agreement with the certified values. This method was compared with those reported in previous sequential injection analysis (SIA) and flow-injection analysis (FIA) systems. The proposed method offers a number of advantages in front the usual AFS applications, which are mainly a higher sampling frequency and a significant reduction in reagent consumption.
Keywords:Multisyringe flow-injection analysis (MSFIA)   Arsenic   Time-based injection   Hydride generation   Atomic fluorescence spectrometry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号