An improved high-performance liquid chromatography method for the determination of homocysteine in human plasma |
| |
Authors: | M. Tröbs T. Renner G. Scherer |
| |
Affiliation: | (1) ABF, Analytisch-biologisches Forschungslabor, Goethestr. 20, 80336 München, Germany |
| |
Abstract: | Summary Elevated plasma homocysteine is, a known risk factor in arteriosclerotic vascular disease. To measure homocysteine in a large number of samples, we have developed a rapid, simple, robust and inexpensive reversed-phase HPLC method for routine analysis. Mercaptopro-pionylglycine was used as the internal standard and an external calibration in plasma was performed. Improvement was achieved by the use of gradient elution (using a sodium acetate buffer and methanol) resulting in a higher number of samples analyzed per day. Plasma samples were reduced with tributylphosphine and the proteins were precipitated with perchloric acid before addition of internal standard. The analytes were derivatized by use of 7-fluorobenzofurazone-4-sulfonic acid ammonium salt. For calibration human plasma was spiked with nine different concentrations of homocysteine (range 2–50 μmol L−1). The inter-assay precision of replicate (n=29) analysis of the concentration of homocysteine in a sample of pooled plasma was 3.0%. The limit of detection, defined as three times the signal-to-noise ratio, was 0.25 μmol L−1. The linearity of the assay was confirmed for a plasma concentration range of 2–2000 μmol L−1. The variation of duplicate analyses of 842 plasma samples was 2.6±1.7%. |
| |
Keywords: | Column liquid chromatography Homocysteine Human plasma Routine analysis |