A Large Volume Injection Procedure for GC-MS Determination of PAHs and PCBs |
| |
Authors: | Z. Večeřa A. Bartošíková J. Sklenská P. Mikuška |
| |
Affiliation: | (1) Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Veveí 97, CZ-61142 Brno, Czech Republic |
| |
Abstract: | A simple on-column injection system for large volume of liquid samples for the GC-MS determination of traces of PAHs and PCBs has been investigated. A deactivated fused silica capillary 20 m × 0.53 mm I.D. and 2 meters of an HP5 column (0.53 mm,1 m film thickness) were used as retention gaps. Injection volumes of 80 L for PAHs and 90 L for PCBs, allow determination of 5–50 ng L–1 PAHs and 11–44 ng L–1 PCBs in hexane solution with an RSD of < 10%. The method has been used for the determination of PCBs and PAHs in soil sample. |
| |
Keywords: | Gas chromatography-mass spectrometry Large volume injection Polycyclic aromatic hydrocarbons Polychlorinated biphenyls Soil samples |
本文献已被 SpringerLink 等数据库收录! |