首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Application of light guides on the PLASMAQUANT-110 ICP spectrometer — Analytical capabilities and analytical performance
Authors:W Quillfeldt
Institution:(1) JENOPTIK Carl Zeiss Jena GmbH, Carl-Zeiss-Strasse 1, O-6900 Jena, Federal Republic of Germany
Abstract:Summary On a high-aperture, high-resolution Echelle spectrometer individual fibre light guides are used to transmit the spectral light from the focal plane to the detectors. For this purpose, 132 light guides are arranged behind an exit slit mask. These light guides transmit the light of the spectral lines of 70 elements. The application of light guides provides numerous advantages for analytical program changes and extensions and enables fast survey analyses. Thoroughly chosen fibre types and precise arrangement of light guides are the prerequisites for a good analytical performance which — beside reproducibilities of lines and background as well as detection limits — is exemplified by the spectral resolution of boron and iron lines as well as of the iron triplet at 310 nm.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏