首页 | 本学科首页   官方微博 | 高级检索  
     


Screen-printed electrodes modified with HRP-zirconium alcoxide film for the development of a biosensor for acetaminophen detection
Authors:Veronica Sima  Cecilia Cristea  Ede Bodoki  Gabriela Duţu  Robert Sāndulescu
Affiliation:(1) Dipartimento di Chimica, Universita' degli Studi della Basilicata, Via N. Sauro 85, 85100 Potenza, Italy
Abstract:
The development of a biosensor based on the immobilization of horseradish peroxidase (HRP) within a zirconium alkoxide-polyetilenimine film onto screen-printed electrodes (SPE) for acetaminophen detection and acetaminophen quantification in pharmaceutical products is described. The biosensor operation mode is based on monitoring the amperometric signal produced by the electrochemical reduction of the enzymatically generated electroactive oxidized species of acetaminophen in the presence of hydrogen peroxide. The enzyme immobilization is performed by retention in a polyethylenimine-zirconium alcoxide porous gel film, a technique that offers good entrapping and a protective environment for the biocomponent due to the hydration properties of the immobilization layer. SPEs have the advantage of being easily mass-produced at low costs with superior characteristics in comparison with classical electrode materials. In this configuration, zirconium alkoxide demonstrates its electrocatalytic activity. The biosensor allows the quantification of acetaminophen with a limit of detection of 6.21×10?8 M and a linear range between 4.35×10?7 M and 4.98×10?6 M. Finally, the biosensor is applied to the quantitative analysis of acetaminophen in Perdolan® tablets.