首页 | 本学科首页   官方微博 | 高级检索  
     


A highly sensitive and stable detection of acetylcholine by HPLC-osmium-horseradish peroxidase redox polymer electrode coated on a gold radial flow ring disk
Authors:Katsunobu Yamamoto  Toshiyuki Chikuma
Affiliation:a Laboratory of Research and Development, Bioelectro Analytical Science Japan Inc., 1-36-6 Oshiage, Sumida-ku, Tokyo 131-0045, Japan
b Laboratories of Molecular Recognition and Natural Information Science, Graduate School of Integrated Science, Yokohama City University, 22-2 Seto, Kanazawa-ku, Yokohama 236-0027, Japan
c Department of Hygienic Chemistry, Showa Pharmaceutical University, Machida 194-8543, Japan
Abstract:To detect a low level of acetylcholine (Ach) output in the dialysate of rat frontal cortex, osmium-peroxidase redox polymer modified gold-ring disk electrode (Os-HRP-RDE) was fabricated in high-performance liquid chromatography (HPLC)-electrochemical detection (ECD). In comparison with platinum electrode, the sensitivity of acetylcholine detection by Os-HRP-RDE was stable during a day, the coefficient of variation was 0.95%, and the decrease of the sensitivity for 1 week was 5.4%. The Os-HRP-RDE could detect lower than 3 fmol of acetylcholine. The present technique, in the absence of acetylcholinesterase (AchE) inhibitor, is useful to facilitate physiological investigations of cholinergic neuronal activity in the brain.
Keywords:Acetylcholine   HPLC   Os-HRP   Au-ring disk   Microdialysis
本文献已被 ScienceDirect 等数据库收录!