Hydrodynamic suppression of the electroosmotic flow in capillary electrophoresis with indirect spectrophotometric detection |
| |
Authors: | S. N. Kalyakin V. V. Sursyakova G. V. Burmakina A. I. Rubailo |
| |
Affiliation: | (1) Institute of Chemistry and Chemical Technology, Siberian Branch, Russian Academy of Sciences, ul. K. Marksa 42, Krasnoyarsk, 660049, Russia;(2) Siberian Federal University, Svobodnyi pr. 79, Krasnoyarsk, 660041, Russia |
| |
Abstract: | A version of capillary electrophoresis with indirect spectrophotometric detection and the hydrodynamic suppression of electroosmotic flow is studied. It is shown that, to improve the reliability of ion identification, one should calculate electrophoretic mobilities of ions or migration times corrected with regard to the electroosmotic flow rate. Correlations between electrophoretic peak areas of ions and their electrophoretic mobilities are derived. In the studied version of capillary electrophoresis, the accuracy of measuring anion concentrations can be improved using the internal standard method. |
| |