首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Voltammetric Determination of Rubeanic Acid
Authors:S V Kovaleva  V P Gladyshev  N M Cheremukhina
Abstract:Procedures for the voltammetric determination of rubeanic acid (RA) at a mercury-film electrode were proposed. They are based on the oxidation of RA at –0.70 to –0.80 V or on the reduction of mercury sulfide, the product of RA oxidation, at –0.82 to –0.85 V (versus a saturated silver–silver chloride electrode) in a 1 M NaOH solution. The oxidation and reduction currents are linear functions of RA concentration in the ranges from 9 × 10–6 to 3 × 10–4 M and from 5 × 10–6 to 3 × 10–4 M, respectively.
Keywords:
本文献已被 SpringerLink 等数据库收录!