A temperature-jump design for conventional NMR probes
Authors:Yushmanov Pavel V  Furó István
Institution:Department of Chemistry, Division of Physical Chemistry and Industrial NMR Center, Royal Institute of Technology, SE-10044 Stockholm, Sweden.
Abstract:The design and performance of a simple probe insert for temperature-jump experiments in conventional NMR probes is described. The insert uses the output from conventional NMR amplifiers for heating conductive aqueous samples with a rate of 30-80 K/s for 200 W radiofrequency power. The observed dependence of the heating rate on sample conductivity is explained by the dominance of dielectric heating. Factors governing the temperature gradient within the sample are discussed.