Amperometric titration of mercury(II) with EDTA, DTPA and TRIEN in the ppm-range
Authors:W. E. van der Linden and J. Dieker
Affiliation:(1) Laboratory for Analytical Chemistry, Amsterdam, The Netherlands
Abstract:Summary Mercury(II) can be titrated at pH 2.5–3 with EDTA, DTPA and TRIEN. The titration is followed amperometrically making use of a rotating gold-electrode at a potential of + 0.25 V vs. S.C.E. Especially the use of DTPA allows the determination of small amounts of mercury(II) (down to 1 mgrg). The standard deviation is 2–3%. A correction for the systematic negative error of approximately 4% in the case of EDTA and 8% in the case of DTPA, mainly caused by the faradayic reaction at the electrode, can easily be made.
Bestimmung von ppm-Mengen Quecksilber(II) durch amperometrische Titration mit ÄDTA, DTPA und TRIEN
Zusammenfassung Quecksilber(II) kann bei pH 2,5–3 mit ÄDTA, DTPA und TRIEN gut titriert werden. Die Titration wird amperometrisch indiziert mit Hilfe einer rotierenden Gold-Elektrode bei einem Potential von + 0,25 V gegen eine gesättigte Kalomelelektrode. Besonders DTPA ermöglicht die Bestimmung von kleinen Mengen Quecksilber(II) (bis etwa 1 mgrg). Die Standardabweichung beträgt 2–3%. Für die systematischen negativen Fehler von etwa 4% bei ÄDTA und 8% bei DTPA, hauptsächlich verursacht durch faradayische Umsetzung an der Elektrode, kann eine einfache Korrektur vorgenommen werden.
Lecture presented at Euroanalysis I Conference, 28. 8. –1. 9. 1972 in Heidelberg, Germany.
Keywords:Best. von Quecksilber(II) mit Ä  DTA, DTPA und TRIEN  Amperometrische Titration