首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and validation of a near infrared method for the analytical control of a pharmaceutical preparation in three steps of the manufacturing process
Authors:M Blanco  J Coello  H Iturriaga  S Maspoch  N Pou
Institution:Departament de Química, Unitat de Química Analítica, Facultat de Ciències, Universitat Autònoma de Barcelona, 08193 Bellaterra, Spain e-mail: iqan8@blues.uab.es, ES
Abstract:A near infrared diffuse reflectance spectroscopy (NIRS) procedure for the quantitative control analysis of the active compound (otilonium bromide) in a pharmaceutical preparation in three steps of the production process (blended product, cores and coated tablets) and a methodology for its validation are proposed. The analytical procedure is composed by two consecutive steps. First, the sample is identified by comparing its spectrum with a second derivative spectral library. If the sample is positively identified, the active compound is quantified by using a previously established partial least squares (PLS) calibration model. The procedure was validated by studying repeatability, intermediate precision, accuracy and linearity. To this end, an adaptation of ICH (International Conference on Harmonisation) validation methodology to an NIR multivariate calibration procedure is proposed. The relative standard error of prediction (RSEP) was ≤ 1% and the suitability of the procedure for control analysis was confirmed by the results obtained analysing new production samples produced over a three-month period.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏