首页 | 本学科首页   官方微博 | 高级检索  
     


Critical assessment of the performance of electronic moisture analyzers for small amounts of environmental samples and biological reference materials
Authors:M Krachler
Affiliation:University of Heidelberg, Institute of Environmental Geochemistry, Germany. krachler@ugc.uni-heidelberg.de
Abstract:Two electronic moisture analyzers were critically evaluated with regard to their suitability for determining moisture in small amounts (< or = 200 mg) of various environmental matrices such as leaves, needles, soil, peat, sediments, and sewage sludge, as well as various biological reference materials. To this end, several homogeneous bulk materials were prepared which were subsequently employed for the development and optimization of all analytical procedures. The key features of the moisture analyzers included a halogen or ceramic heater and an integrated balance with a resolution of 0.1 mg, which is an essential prerequisite for obtaining precise results. Oven drying of the bulk materials in a conventional oven at 105 degrees C until constant mass served as reference method. A heating temperature of 65degrees C was found to provide accurate and precise results for almost all matrices investigated. To further improve the accuracy and precision, other critical parameters such as handling of sample pans, standby temperature, and measurement delay were optimized. Because of its ponderous heating behavior, the performance of the ceramic radiator was inferior to that of the halogen heater, which produced moisture results comparable to those obtained by oven drying. The developed drying procedures were successfully applied to the fast moisture analysis (1.4-6.3 min) of certified biological reference materials of similar provenance to the investigated the bulk materials. Moisture results for 200 mg aliquots ranged from 1.4 to 7.8% and good agreement was obtained between the recommended drying procedure for the reference materials and the electronic moisture analyzers with absolute uncertainties amounting to 0.1% and 0.2-0.3%, respectively.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号