首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of Ochratoxin A in wine at sub ng/mL levels by solid-phase microextraction coupled to liquid chromatography with fluorescence detection
Authors:Aresta Antonella  Vatinno Rosa  Palmisano Francesco  Zambonin Carlo G
Affiliation:Dipartimento di Chimica, Università degli Studi di Bari, Via E. Orabona 4, 70126 Bari, Italy.
Abstract:Solid-phase microextraction (SPME), using a polydimethylsiloxane/divinylbenzene (PDMS/DVB) fiber, interfaced with liquid chromatography-fluorescence detection (LC-FD) has been applied to the determination of Ochratoxin A (OTA) in wine samples. Compared to the most widely adopted extraction/clean-up procedure based on immunoaffinity columns (IAC), the solventless extraction is simpler and cost-effective, requiring the simple immersion of the fiber in diluted wine samples. Furthermore, a fast LC separation is achieved under isocratic conditions. The linear range investigated in wine was 0.25-8 ng/mL; at fortification levels of 0.5 and 2 ng/mL, within-day intra-laboratory precision (repeatability) values, expressed as RSD%, were 5.9 and 5.1, respectively, whereas between days (n = 4) precision was 8.5 and 7.1%, respectively. The limit of detection (LOD) at a signal-to-noise (S/N) ratio of 3 was 0.07 ng/mL; the limit of quantification (LOQ) calculated at S/N = 10 was 0.22 ng/mL, well below the European regulatory level of 2 ng/mL. The potential of the method has been demonstrated by the analysis of a number of different wine samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号