Determination of ethanol in alcoholic beverages by high-performance liquid chromatography-flame ionization detection using pure water as mobile phase
Authors:Yarita Takashi  Nakajima Ryoji  Otsuka Satoko  Ihara Toshihide A  Takatsu Akiko  Shibukawa Masami
Abstract:A method for the determination of ethanol in alcoholic beverages by high-performance liquid chromatography-flame ionization detection (HPLC-FID) was developed. An FID system could be directly connected to an HPLC system using pure water as a mobile phase. In a durability test using triacontylsilyl (C30)-silica gel stationary phase for 96 h, no significant change in the retention time of four alcohol compounds was observed. So the HPLC separation of alcoholic beverages was carried out on the C30-silica gel stationary phase. On application to the analysis of six kinds of alcoholic beverages, ethanol could be determined accurately by the proposed method.
Keywords:
本文献已被 PubMed 等数据库收录!