Determination of inorganic anions in natural water by microcolumn ion chromatography with on-column enrichment. |
| |
Authors: | Lee Wah Lim Ji-Ye Jin Toyohide Takeuchi |
| |
Affiliation: | Department of Chemistry, Faculty of Engineering, Gifu University, 1-1 Yanagido, Gifu 501-1193, Japan. |
| |
Abstract: | An on-column enrichment method was developed for the rapid determination of inorganic anions in natural water. The system was assembled from a syringe pump, a six-port switching valve with a sample-enrichment loop, a separation column and a UV detector. The enrichment efficiency of the system was tested by using inorganic anions as samples. The limits of detection were between 0.6 and 7.7 microg/L. The system was applied to the determination of anions in river and pond-water samples. |
| |
Keywords: | |
|
|