Determination of ascorbic acid by use of a flow-through solid phase UV spectrophotometric system |
| |
Authors: | A Molina-Díaz A Ruiz-Medina and M L Fernández-de Córdova |
| |
Institution: | (1) Department of Physical and Analytical Chemistry, Faculty of Experimental Sciences, University of Jaén, E-23071 Jaén, Spain e-mail: amolina@ujaen.es, ES |
| |
Abstract: | A continuous flow-through solid phase spectrophotometric system was developed for the determination of ascorbic acid based
on the measurement of its intrinsic absorbance in the UV region when retained on a 1 mm Sephadex QAE A-25 anion exchanger
gel layer which is placed into an appropriate quartz flow-through cell, the absorbance exhibited by this solid phase being
monitored at 267 nm. A monochannel manifold was used, the sample (300, 600 or 1000 μL) being injected into the carrier solution
(acetate buffer). This solution also elutes the analyte after developing the analytical signal, and regenerates the resin
layer which, therefore, remains ready for the next sample. The linear dynamic range and other analytical parameters vary according
to the sample volume injected. Three calibration lines were established for 300, 600 and 1000 μL sample volume, which ranged
from 1.0 to 20.0, 0.5 to 10.0 and 0.2 to 6.0 μg mL–1, respectively. The detection limits were 0.04 (300 μL), 0.03 (600 μL) and 0.02 μg mL–1 (1000 μL), the sampling rates 28, 24 and 21 h–1, and the RSDs (n = 10) 0.87%, 1.08% and 0.90%, respectively. The amount of ascorbic acid in various samples (pharmaceuticals, sweets and urine)
were successfully determined with this method.