Affiliation:(1) Adelaide Laboratory, CSIRO Land and Water, PMB 2, 5064 Glen Osmond, SA, Australia
Abstract:Summary On-line solid-phase extraction (SPE) coupled with reversed-phase liquid chromatography and UV detection at 254 nm has been used for the determination of trace-level polycyclic aromatic hydrocarbons (PAH) in soil extracts. Five commercially available adsorbents (C8, C18, PLRP-S, PRP-1, and Bond-Elut Env) were evaluated. Results showed that recovery of the PAH decreased with increasing molecular weight, because of their poorer solubility. Recovery of high-molecular-weight PAH was significantly improved by addition of 10% (v/v) acetonitrile to the sample before loading of the SPE adsorbent. PAH recovery ranged from 64.0 to 108% when a 50 mL sample spiked with 1 μg L−1 was applied to these adsorbents. Determination of PAH was possible with detection limits below 0.05 μg L−1, which corresponds to 0.2 μg kg−1 soil. The method was successfully used to determine PAH in soil extracts.
Keywords:Column liquid chromatography  On-line solid-phase extraction  Soil samples  Trace PAHs