Abstract: | On-line coupling continuous-flow liquid membrane extraction (CFLME) with HPLC, a novel automatic system was developed for the determination of sulfonylurea herbicides in water. After an automatic trace-enrichment process by CFLME, which is the combination of continuous flow liquid-liquid extraction and support liquid membrane (SLM) extraction, the target analytes were concentrated in 50 microl of 0.2 M Na2CO3-NaHCO3 (pH 10.0) buffer. The concentrated sample solutions were injected directly onto a C18 analytical column with a valve, and detected at 240 nm with a diode array detector. Metsulfuron methyl (MSM), and DPX-A 7881 were baseline separated with a mobile phase consisting of methanol and 67 mM KH2PO4-Na2HPO4 (pH 5.91) buffer (45+55, v+v) at a flow-rate of 1.0 ml min(-1). With an enrichment time of 10 min and enrichment sample volume of 20 ml, the enrichment factors and detection limits are 100 and 0.05 microg l(-1) for MSM, and 96 and 0.1 microg l(-1) for DPX-A 7881, respectively. The linear range and precision (RSD) are 0.1-50 microg l(-1) and 7.0% for MSM, and 0.2-50 microg l(-1) and 9.2% for DPX-A 7881, respectively. This proposed method was applied to determine MSM and DPX-A 7881 in seawater, tap water, and bottled mineral water with spiked recoveries in the range of 83-95% for MSM and 88-100% for DPX-A 7881, respectively. |
| |