首页 | 本学科首页   官方微博 | 高级检索  
     


Studies on an antibody-coated tube method of aldosterone
Authors:S Ishihara  H Moriuma  C Mishima  M Honda  M Tomonobu  K Kanao  N Usami
Affiliation:Radiosotope Department, Sumitomo Hospital, Osaka-shi, Japan.
Abstract:We studied on the basic performance of a non-extraction coated tube RIA kit, DPC Aldosterone kit (Nippon DPC Corporation). 1. The average binding CVs of the calibrators (25 to 1,200 pg/ml) were stable in a range from 1.4 to 1.9%. 2. The minimum sensitivity was 17 pg/ml from 2SD method of 0 calibrator. 3. The intra-assay reproducibility test showed an average CV as 9.0%, and the inter-assay test showed as 8.8%. 4. The recovery ratios of a recovery test were 85.4 to 121.6%. 5. The correlation test with another maker's kit showed a correlation curve as y = 0.970 x + 19.68 and a correlation coefficiency as r = 0.970. 6. The normal range, the correlation between fasting recumbency and daytime sitting, and the correlation between walking and resting, showed similar results as former reports.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏