Interpolation in the standard additions method
Authors:JM Andrade  J Terán-BaamondeRM Soto-Ferreiro  A Carlosena
Institution:Department of Analytical Chemistry, Faculty of Sciences, University of A Coruña, Campus da Zapateira s/n, E-15008 A Coruña, Spain
Abstract:The standard additions method (SAM) has traditionally been performed by using extrapolation. This practice is suboptimal because predictions are affected by even slight departures of calibration points from a straight line. Despite this, most textbooks and papers in analytical chemistry still refer exclusively to extrapolation. In contrast, the use of interpolation is recommended in this paper as a way to get predictions on the central part of the regression line and thus minimize the bias in the prediction and the variance associated with the analytical result. Several scenarios were studied, with concentration errors simulated in different calibration solutions. It was found that translational effects due to variations at the central part of the calibration caused the lowest disturbances on the predicted concentrations. The differences between the interpolated and extrapolated predictions can be as large as ±30%. The confidence interval associated with the extrapolation result is wider than that due to interpolation by as much as 100%. It is shown that commonly used equations underestimate the correct confidence intervals. Both, absence of bias and improved precision, are of relevance in quality assurance, method validation and error propagation.
Keywords:Standard additions method  Interpolation  Extrapolation  Method validation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏