a Dionex Corporation, 1228 Titan Way, P.O. Box 3603, Sunnyvale, CA 94088-3603, USA
b Dionex GmbH, Am Wörtzgarten 10, D-65510 Idstein, Germany
Abstract:
An ion chromatography method was developed to determine trace anionic contamination in matrices that have a high concentration of nitrate ion. Contaminant anions of interest were separated on an IonPac AS15 high-capacity anion-exchange column and detected by suppressed conductivity detection. An EG40 eluent generator was used to prepare high-purity and carbonate-free potassium hydroxide. Using the EG40, performance at trace levels was enhanced because background conductivity decreased and retention time reproducibility improved. Trace anionic contamination from the mobile phase was minimized when using the eluent generator compared to using conventionally prepared sodium hydroxide eluents. The signal-to-noise ratio was also improved with the use of a temperature controlled conductivity cell and chromatography hardware in the microbore (2-mm) format. The eluent concentration was optimized to separate the contaminant anions from the excess of the nitrate matrix ions. The procedure was demonstrated for a solution of reagent-grade sodium nitrate and high-purity 0.7% nitric acid. Method detection limits for chloride, sulfate and phosphate of 150 μg/l and lower were achieved.