Producing SI-traceable reference values for Cd, Cr and Pb amount contents in polyethylene samples from the Polymer Elemental Reference Material (PERM) project using isotope dilution mass spectrometry |
| |
Authors: | J. Vogl D. Liesegang M. Ostermann J. Diemer M. Berglund C. R. Quétel P. D. P. Taylor K. G. Heumann |
| |
Affiliation: | (1) European Commission, Joint Research Center, Institute for Reference Materials and Measurements, Retieseweg, 2440 Geel, Belgium e-mail: christophe quetel@irmm.jrc.be Tel.: +32-14-571-658 Fax: +32-14-571-863, BE;(2) Johannes Gutenberg-University of Mainz, Institute of Inorganic and Analytical Chemistry, Duesbergweg 10–14, 55099 Mainz, Germany, DE |
| |
Abstract: | The present paper describes the certification of the amount content of Cd, Cr and Pb in two different polyethylene materials within the third phase of the Polyethylene Elemental Reference Material (PERM) project. The analytical procedure to establish the reference values for Cd, Cr and Pb amount contents in these materials is based on isotope dilution mass spectrometry used as a primary method of measurement. Cd and Pb were measured with inductively coupled plasma-mass spectrometry and Cr with positive thermal ionization-mass spectrometry. The decomposition of the polymer matrix was carried out using a high pressure asher. Reference values for amount content, traceable to the SI-system, have been obtained for these three elements in both of the polyethylene samples of PERM. For each of the certified amount content values an uncertainty budget was calculated using the method of propagation of uncertainties according to ISO and EURACHEM guidelines. The measurement procedures, as well as the uncertainty calculations, are described for all three elements. In order to keep the whole certification process as transparent as possible, the preparations of various reagents and materials as well as the sample treatment and blending are described in detail. The mass spectrometry measurements and the data treatment are also explained carefully. The various sources of uncertainty present in the procedure are displayed in the uncertainty budgets. The obtained combined uncertainties for the amount content values were less than 2% relative (k=1) for all investigated elements. The amount contents were in the μmol/kg range, corresponding to mg/kg levels. Received: 21 October 1999 / Accepted: 29 January 2000 |
| |
Keywords: | PERM ICP-MS PTI-MS IDMS Uncertainty budget Certification Polyethylene |
本文献已被 SpringerLink 等数据库收录! |
|