Validation of a high-performance liquid chromatographic method with fluorescence detection for the simultaneous determination of tetracyclines residues in bovine milk
Authors:Spisso Bernardete Ferraz  de Oliveira e Jesus Ana Luiza  de Araújo Marcus Antônio Gonçalves  Monteiro Mychelle Alves
Institution:National Institute for Quality Control in Health, Oswaldo Cruz Foundation, Av. Brasil 4365, Rio de Janeiro, Brazil
Abstract:A high-performance liquid chromatography-fluorescence detection method was optimized and validated to determine tetracyclines residues in bovine milk. Post-column derivatization using metal complexation in non-aqueous reagent increased the fluorescence of chelates by a factor up to 2.54 compared to water (signal-to-noise ratio enhancement). Overall recoveries ranged from 61 to 115%, with RSDr from 5 to 15% (n = 54). Detection limits ranged from 5 to 35 μg kg−1. Limits of quantification were established at 50 μg kg−1. Decision limits (CCα) were 109, 108 and 124 μg kg−1 and detection capabilities (CCβ) 119, 117 and 161 μg kg−1 for oxytetracycline, tetracycline and chlortetracycline, respectively. The method was applied successfully in a national monitoring program.
Keywords:Validation  Milk  Residues  Tetracyclines  Antibiotics