Anticosmic-shielded ultralow-background germanium detector systems for analysis of bulk environmental samples |
| |
Authors: | J H Reeves R J Arthur |
| |
Institution: | (1) Pacific Northwest Laboratory, Richland, Washington, USA |
| |
Abstract: | A low-level gamma-ray counting system has been developed which reduces system background, relative to other typical low-background systems, by a factor of ten in the energy region below one MeV, and by as much as a factor of forty at higher energies. This germanium-diode gamma-ray spectrometer was constructed for a modest investment above that required for a conventional germanium detector. The techniques involved use: (1) materials of known radiopurity to surround the diode, (2) an active external anticosmic shield to reduce the background continuum due to interactions of cosmic particles with the detector and passive lead shielding, and (3) nitrogen exhausted from the cryogenic dewar to minimize the introduction of ubiquitous radon decay nuclei into the sample counting chamber. A novel method for handling samples prior to counting is presented. Also; some of the difficulties encountered in calibrating a system intended for bulk samples are discussed.Operated for the U.S. Department of Energy by Battelle Memorial Institute under Contract DE-AC06-76RLO 1830. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|