Poly(glycidylmethacrylate-co-vinyl ferrocene)-grafted iron oxide nanoparticles as an electron transfer mediator for amperometric phenol detection |
| |
Authors: | Emre Çevik Mehmet Şenel Abdulhadi Baykal M Fatih Abasıyanık |
| |
Institution: | 1. Department of Genetics and Bioengineering, Faculty of Engineering, Fatih University, B.Cekmece, Istanbul 34500, Turkey;2. Department of Chemistry, Faculty of Arts and Sciences, Fatih University, Buyukcekmece Kampusu, B.Cekmece, Istanbul 34500, Turkey |
| |
Abstract: | An amperometric phenol biosensor was constructed by using poly(glycidylmethacrylate-co-vinyl ferrocene) grafted iron oxide nanoparticles for detection of different phenolic compounds (catechol, aminophenol, phenol, p-cresol, pyrogallol). The poly(glycidylmethacrylate-co-vinyl ferrocene) and nanoparticles were characterized by Fourier transform infrared (FT-IR), X-ray diffraction (XRD) and transmission electron microscopy (TEM). The copolymer grafted iron oxide nanoparticles and Horseradish peroxidase (HRP) were covalently attached on gold (Au) electrode surface. The effect of pH, temperature and characteristic features such as; reusability and storage stability were studied. The electrode showed good response time within ~3 s. The electrocatalytic response showed a linear dependence on the phenolic compounds concentration ranging from 0.5 to 17.0 mM. |
| |
Keywords: | Horseradish peroxidase Magnetic nanoparticles Ferrocene biosensor Phenolic compounds |
本文献已被 ScienceDirect 等数据库收录! |
|