Fast analysis of terbutaline in pharmaceuticals using multi-walled nanotubes modified electrodes from recordable compact disc |
| |
Authors: | Fabiana S. Felix,Daniela Daniel,Jivaldo R. Matos,Claudimir Lucio do Lago,Lú cio Angnes |
| |
Affiliation: | 1. Universidade de São Paulo, Instituto de Química, São Paulo, 05508-000, Brazil;2. Agilent Technologies, Alameda Araguaia, 1142, CEP 06455-000, Barueri, SP, Brazil |
| |
Abstract: | In this study, homemade disposable gold electrodes made from recordable compact disks were modified with carbon nanotubes for amperometric quantification of terbutaline sulfate in pharmaceutical products. A flow cell using an impingent jet of solution on the electrode surface was build and used for amperometric detection, and a series of experiments were carried out to find the best experimental conditions for the new electrode in a specially designed cell. A linear response for terbutaline was obtained in the range from 3.0 × 10−6 to 5.0 × 10−4 mol L−1 (at 0.63 V vs. Ag/AgCl). The limits of detection and quantification were calculated as 5.8 × 10−7 mol L−1 (S/N = 3) and 1.9 × 10−6 mol L−1 (S/N = 10), respectively. A frequency of 30 injections h−1 was attained. The proposed method was successfully applied to the analyses of commercial syrup samples, and all results were in good agreement with those obtained by using high performance liquid chromatography and capillary electrophoresis-tandem mass spectrometry. |
| |
Keywords: | Multi-walled nanotubes Gold electrodes from recordable compact disc Terbutaline Amperometry Impingent flowing system |
本文献已被 ScienceDirect 等数据库收录! |
|