Abstract: | The application of pulsed cavity ring-down spectroscopy (CRDS) was demonstrated for the measurement of nitrogen dioxide (NO2) in automotive exhaust gas. The transition of the ν 3 vibrational band assigned to the antisymmetric stretching mode of NO2 was probed with a thermoelectrically cooled, pulsed, mid-infrared, distributed feedback, quantum cascade laser (QCL) at 6.13 μm. The measurement of NO2 in the exhaust gas from two diesel vehicles equipped with different aftertreatment devices was demonstrated using a CRDS-based NO2 sensor, which employs a HEPA filter and a membrane gas dryer to remove interference from water as well as particulates in the exhaust gas. Stable and sensitive measurement of NO2 in the exhaust gas was achieved for more than 30 minutes with a time resolution of 1 s. |