首页 | 本学科首页   官方微博 | 高级检索  
     


Correction of the interference by mixing the reagents in the bioluminescent determination of ATP in environmental samples by an interfaced PC photometer and a robust regression procedure
Authors:M. Mecozzi  Marina Amici  Giovanni Visco
Affiliation:(1) ICRAM, Istituto Centrale per la Ricerca Scientifica Applicata al Mare, via L. Respighi n. 5, I-00197 Rome, Italy, IT;(2) Università degli Studi La Sapienza di Roma, piazzale Aldo Moro n. 5, Rome, Italy, IT
Abstract: An automated procedure for correcting the interference due to mixing the reagents in the determination of ATP by a bioluminescent technique is described. The measurements were performed by an in-house interfaced PC photometer. An automated procedure based on a robust regression method and outliers detection was applied to exclude the time of the mixing of the reagents from the global time of reaction. In fact the mixing of the reagents is responsible of the poor analytical precision (RSD%>25) and low sensitivity (limit of detection≈2×10-8 mol/l). By this procedure, the precision (RSD%<7) and sensitivity (limit of detection≈8×10-10 mol/l) were improved. The method was applied to the determination of ATP in marine sediments and good recovery within 90–108% was obtained. For non-polluted samples the measurement can be carried out by direct comparison with standard solutions. The standard addition procedure is recommended for polluted samples. Received: 25 April 1996/Revised: 28 June 1996/Accepted: 3 July 1996
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号