An evaluation of a new neutron tube (Philips Type 18604) for total body in vivo activation analysis |
| |
Authors: | J K Haywood I Robertson K Boddy |
| |
Institution: | (1) Health Physics and Nuclear Medicine Unit, Scottish Universites Research and Reactor Centre, East Kilbride, G75 OQU Glasgow, Scotland |
| |
Abstract: | A 14 MeV neutron generator system with a new sealed tube (Philips 18604) was evaluated for total body in vivo neutron activation
analysis (TBIVNAA). The neutron output, angular distribution, rise time, reproducibility of short irradiations and leakage
of tritium were investigated. The 18604 tube meets all the requirements of TBIVNAA within its normal operating specification. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|