首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Discrepancy in infrared measurement results of carbon monoxide in nitrogen mixtures due to variations of the 13C/12C isotope ratio
Authors:Gerard Nieuwenkamp  Adriaan M H van der Veen
Institution:(1) Nederlands Meetinstituut, Thijsseweg 11, 2629 JA Delft, The Netherlands
Abstract:Significant errors in the non-dispersive infrared (NDIR) analyses of carbon monoxide (CO) can be made when the 13C/12C isotope ratio in the sample and the calibrant differ significantly. This paper shows that variations in the 13C/12C isotope ratio of 5×10−2 mol/mol CO in nitrogen mixtures on three different NDIR CO analysers may lead to serious deviations in the instrument response, whereas the instrument response using GC-TCD is unaffected. The observed deviations in the assigned amount-of-substance fraction CO for a 13C depleted mixture vary from +2 to −5% relative to the gravimetric amount-of-substance fraction for different NDIR analysers. A GC-MS method has been developed to perform a pre-screening of the isotopic composition of CO in nitrogen mixtures. This method proved to be an adequate tool to measure differences in the 13C/12C ratio. Based on the GC-MS results a suitable measurement technique can be selected, or information about a possible error in NDIR analysis can be given to the producer or user of the calibration gas mixture. Presented at 3. International Gas Analysis Symposium, October 6–8, 2004, in Amsterdam
Keywords:NDIR                  13C  Isotope ratio
本文献已被 SpringerLink 等数据库收录!