Simplified method for polarization-sensitive optical coherence tomography
Authors:Roth J E  Kozak J A  Yazdanfar S  Rollins A M  Izatt J A
Abstract:We report a method for extracting the birefringence properties of biological samples with micrometer-scale resolution in three dimensions, using a new form of polarization-sensitive optical coherence tomography. The method measures net retardance, net fast axis, and total reflectivity as a function of depth into the sample. Polarization sensing is accomplished by illumination of the sample with at least three separate polarization states during consecutive acquisitions of the same pixel, A scan, or B scan. The method can be implemented by use of non-polarization-maintaining fiber and a single detector. In a calibration test of the system, net retardance was measured with an average error of 7.5 degrees (standard deviation 2.2 degrees ) over the retardance range 0 degrees to 180 degrees , and a fast axis with average error of 4.8 degrees over the range 0 degrees to 180 degrees .