首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Label‐Free Detection of Bisphenol A Using a Potentiometric Immunosensor
Authors:Ming‐Hua Piao  Hui‐Bog Noh  Md?Aminur Rahman  Mi‐Sook Won  Yoon‐Bo Shim
Institution:1. Department of Chemistry and Center for Innovative Biophysio Sensor Technology, Pusan National University, Busan 609‐735, S. Korea;2. Korea Basic Science Institute, Busan 609‐735, S. Korea
Abstract:A potentiometric immunosensor for the label‐free detection of bisphenol A (BPA) was developed by covalently immobilizing a polyclonal antibody (PAb) onto the surface of a carboxylated poly(vinyl chloride) (PVC‐COOH) membrane. The immunosensor was characterized using scanning electron microscopy (SEM), X‐ray photoelectron spectroscopy (XPS), and electrochemical impedance spectroscopy. The immobilization of PAb and its interaction with antigen (Ag) and BPA were also confirmed by quartz crystal microbalance (QCM) studies. Experimental parameters affecting the immuno‐interaction between PAb and its Ag or BPA were examined in terms of pH, antibody concentration, and temperature. The immunosensor showed a specific recognition of BPA with less interference than 4.8% from other common phenolic compounds. A calibration plot was obtained between 1.0 and 30.0 ng/mL and the detection limit was determined to be 0.6 ng/mL. The proposed immunosensor was applied for a real water sample spiked with BPA and the recoveries were in the range between 102.5 and 105.7%.
Keywords:Polyclonal antibody  Label‐free detection  Potentiometric method  Immunosensor  Bisphenol A
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号