Determination of the purity of potassium iodate by constant-current coulometry |
| |
Authors: | L. Ma |
| |
Affiliation: | (1) National Research Center For Certified Reference Material, No.18 Bei San Huan Dong Lu, Chaoyang Qu, Beijing 100013, P.R. China e-mail: Mald@nrccrm.com.cn Tel.: 0086-010-64294060 Fax: 0086-010-64278404, CN |
| |
Abstract: | This paper describes a method of determining the purity of potassium iodate by constant-current coulometry. The determination can be divided into two steps. First, a Na2S2O3 solution is prepared and its reductive value is determined. Second, the purity value of an oxidimetric quantity of potassium iodate is determined. This paper discusses the conditions of the reaction process and evaluates type A and B standard uncertainty of this method. The expanded relative uncertainty of this method is 0.02% (k=2). Received: 15 June 2000 Accepted: 27 December 2001 |
| |
Keywords: | Potassium iodate Constant-current coulometry Oxidimetric quantity |
本文献已被 SpringerLink 等数据库收录! |
|