首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of an electrical aerosol detector (EAD) for the aerosol integral parameter measurement
Authors:Lin Li  Da-Ren Chen  Perng-Jy Tsai
Affiliation:1. Department of Energy, Environmental and Chemical Engineering, Washington University in St. Louis, One Brookings Drive, Box 1180, St. Louis, MS 63130, USA;2. Department of Environmental and Occupational Health, Medical College, National Cheng Kung University, 138 Sheng-Li Road, 70428 Tainan, Taiwan
Abstract:The performance of an electrical aerosol detector (EAD; TSI Model 3070A) was experimentally evaluated for measuring the integral parameters of particles (i.e., total length concentration of particles, and the total surface area concentrations of particles deposited in a human lung). The EAD consists of a unipolar diffusion charger with an ion trap, and aerosol electrometer. We first evaluated the performance of the EAD charger. Both polydisperse and monodisperse particles of Ag, NaCl, and oleic acid (with the dielectric constants of infinite, 6.1 and 2.5) were then generated to evaluate the particle material effect on the EAD readout.
Keywords:
本文献已被 ScienceDirect 等数据库收录!